800 Dosino - General

The 800 Dosino is a flexible dosing drive which, as an MSB device, can be used with a host of Metrohm main devices. Together with the intelligent 807 Dosing unit, it is well suited for both simple dosing and for automation and liquid handling tasks, such as sample transfer or pipetting.

Note MagIC Net only supports the use of intelligent 807 Dosing units!

Configuration MSB Device

An MSB device is not entered in the device table. In the configuration it is only listed under the MSB connections of the main device to which it is connected (e.g. 850 Professional IC or 858 Professional Sample Processor).

800 Dosino - Method

An 800 Dosino (as is the case with all MSB devices) cannot be added separately to a method. It is added to the list of devices by adding the main device to which it is connected to the method.

If the method is created without connected devices, the MSB device can be added separately. In this case, the MSB connector on the main device must be entered manually.

In the device group, the 800 Dosino behaves like a stand-alone device. It can be selected and displayed graphically in the upper area of the device window.

800 Dosino - Start parameters

Tab: Method ? Devices ? Dosino Solution

Select the solution for the Dosino that should be available when the hardware is started with [Start HW]. A check is carried out in the method run to ensure that the correct solution has been put on the dosing device. When the method is started, the working life and GLP test interval are checked for the selected solution.

Selection not defined | Solutions from the solution table Default value not defined not defined

The solution is not checked.

Solutions from the solution table

Only \solutions, i.e. solutions that are assigned to an 807 Dosing unit, can be selected. Connector

Shows the main device and the MSB connector which the 800 Dosino is on.

800 Dosino - Commands

Overview of the time program commands available for the 800 Dosino:

Dosing

Dialog window: Method ? Time program ? Edit ? 'Device' - Dosing

The device-dependent time program command Dosing (command with feedback) doses the specified volume via the defined port. There is no automatic filling beforehand or afterwards.

Parameters

Port

Port via which dosing takes place. Formula input possible.

Range Default value Volume

1 ... 4 1

Volume to be transported. Formula input possible.

Range Default value Dosing rate

0.0000 ... 99999.9000 mL 1.0000 mL

Speed at which discharge takes place. The maximum rate depends on the cylinder volume of the buret used. Formula input possible.

Cylinder volume 2 mL

Range Default value Selection Range Default value Selection Range Default value Selection Range Default value Selection Range Default value Selection

0.01 ... 6.00 mL/min 1.00 mL/min maximum

Cylinder volume 5 mL

0.01 ... 16.00 mL/min 1.00 mL/min maximum

Cylinder volume 10 mL

0.01 ... 33.00 mL/min 1.00 mL/min maximum

Cylinder volume 20 mL

0.01 ... 66.00 mL/min 1.00 mL/min maximum

Cylinder volume 50 mL

0.01 ... 166.00 mL/min 1.00 mL/min maximum

Filling rate

Speed at which filling takes place. The maximum rate depends on the cylinder volume of the buret used. Formula input possible.

Cylinder volume 2 mL

Range Default value Selection Range Default value Selection Range Default value Selection Range Default value Selection Range Default value Selection

0.01 ... 6.00 mL/min 1.00 mL/min maximum

Cylinder volume 5 mL

0.01 ... 16.00 mL/min 1.00 mL/min maximum

Cylinder volume 10 mL

0.01 ... 33.00 mL/min 1.00 mL/min maximum

Cylinder volume 20 mL

0.01 ... 66.00 mL/min 1.00 mL/min maximum

Cylinder volume 50 mL

0.01 ... 166.00 mL/min 1.00 mL/min maximum

Comment

Optional comment on the time program command.

Input

64 characters

Generated variables

The following type 'TP.command number.Variable name' variables are generated by the command: Variable name Description CONC Data type Concentration of the dosed solution as long as one Number DVOL has been assigned to the Dosino. Dosed volume [mL] Number Aspirate

Dialog window: Method ? Time program ? Edit ? 'Device' - Aspirate

The device-dependent time program command Aspirate (command with feedback) draws in the specified volume via the defined port. As with the command Dosing, the cylinder is not filled automatically beforehand or afterwards. It should be possible to achieve the volume to be aspirated with a single piston stroke.

Parameters

Port

Port via which aspiration takes place. Formula input possible.

Range Default value Volume

1 ... 4 1

Volume to be transported. Formula input possible.

Range Default value Rate

0.0000 ... 50.0000 mL 1.0000 mL

The maximum rate depends on the cylinder volume of the buret used. Formula input possible.

Cylinder volume 2 mL

Range Selection Default value Range Selection Default value

0.01 ... 6.00 mL/min maximum maximum

Cylinder volume 5 mL

0.01 ... 16.00 mL/min maximum maximum

百度搜索“77cn”或“免费范文网”即可找到本站免费阅读全部范文。收藏本站方便下次阅读,免费范文网,提供经典小说教育文库瑞士万通离子色谱仪800 Dosino进样器说明书--英文在线全文阅读。

瑞士万通离子色谱仪800 Dosino进样器说明书--英文.doc 将本文的Word文档下载到电脑,方便复制、编辑、收藏和打印 下载失败或者文档不完整,请联系客服人员解决!
本文链接:https://www.77cn.com.cn/wenku/jiaoyu/1050501.html(转载请注明文章来源)
Copyright © 2008-2022 免费范文网 版权所有
声明 :本网站尊重并保护知识产权,根据《信息网络传播权保护条例》,如果我们转载的作品侵犯了您的权利,请在一个月内通知我们,我们会及时删除。
客服QQ: 邮箱:tiandhx2@hotmail.com
苏ICP备16052595号-18
× 注册会员免费下载(下载后可以自由复制和排版)
注册会员下载
全站内容免费自由复制
注册会员下载
全站内容免费自由复制
注:下载文档有可能“只有目录或者内容不全”等情况,请下载之前注意辨别,如果您已付费且无法下载或内容有问题,请联系我们协助你处理。
微信: QQ: